A Notified Body’s Perspective on Post-Market PhaseDownload or print this article (PDF Format)As economic belts tighten around the world the medical device industry is one of the hardest-hit, as one of the most competitive industries. While competition has increased, medical device recalls around the world have also increased in recent years. European authorities recalled 434 devices in 2008, 611 devices in 2009, and 748 devices in 2010.1 A similar trend appeared in the United States in which the FDA recalled 18 devices in 2008, 31 devices in 2009, and 45 devices in 2010. Recalls may seem alarming to the public, but a recall is an integral part of a manufacturer’s post-market surveillance system to answer to public health problems and protect patients from any new occurrence of an incident. In addition, a recall could not only be a failure of the product but also a misuse by doctors for various reasons like inappropriate or incomplete instructions for use. A notified body, in essence, cannot provide any consulting assistance about post-market surveillance systems. However, choosing a notified body that is dedicated to its regulatory duties in assessing its clients’ QMS can help a manufacturer to be sure that its own provisions on this matter are likely to ensure he efficiently manages the available product data after the marketing of the product. The Directives themselves provide little guidance for a manufacturer’s duties for conducting post-market surveillance except that they must include a surveillance system in their quality management system. To begin, the manufacturer’s post-market surveillance system must allow the manufacturer to collect data from potentially each user over the lifetime of the device. The manufacturer will then use this data to continually analyze the device’s risk to benefit ratio. Collecting Information - Setting Up a Post-Market Surveillance SystemThe data that manufacturers must collect is:
The European Commission recently released a new MEDDEV guidance document on PMCF studies and their importance to the manufacturer’s vigilance procedures.3 It provides guidance for manufacturers and notified bodies assessing the system for “identifying and investigating residual risks associated with the use of medical devices placed on the market.”4 Using the device’s risk management file, manufacturers can decide to conduct a PMCF study to monitor any residual risks and continually examine the device’s risk to benefit ratio. According to the MEDDEV guidance document, some circumstances that arise that justify a PMCF include significant changes or innovation to the device and where the device would be implanted.5 A PMCF study should be outlined in the manufacturer’s post-market surveillance information for the specific device. According to the guidance document, the study outline should include:
Generally, the study should be designed to answer the research questions and address the study’s objectives. The study may conform to different methodologies and it should describe the methodology7 to ensure the manufacturer can draw conclusions from the study. Note that the evaluation by the notified body or a competent authority will be easier if the methodology is clearly described. While manufacturers have the duty to collect data, they can delegate the complaint-collecting task to local distributors. Because manufacturers have this duty, however, they must make sure to include this task in their contracts and enforce it. Essentially, a manufacturer should be able to use the post-market surveillance system they design to collect data, analyze the data, and act on the data they receive. Analyzing and Acting on Data Received by the Post-Market Surveillance SystemCase of the Data Related to an Incident The post-market surveillance system must have a method for manufacturers to analyze the data related to incidents it collects from its PMCF studies and complaint collecting. For medical devices sold in Europe, the criteria and procedure manufacturers use to report incidents is found in MEDDEV 2.12-1 rev 6, Guidelines on a Medical Devices Vigilance System. This guidance document outlines the procedure manufacturers should use to determine if an event should be reported and the duties of each actor in the vigilance process. The guidance document first specifies the elements of what a reportable incident is for a competent authority. According to reporting criteria, when a manufacturer receives a complaint that an event that is linked to their device occurs and a patient or user dies or experiences a serious deterioration in health, the manufacturer must report the incident.8 These incidents must be reported to the relevant competent authorities for the countries in which the manufacturer sells the device in a field incident report. Depending on the severity of the event, the manufacturer may need to report the incident sooner rather than later. The manufacturer’s notified body may be informed too. The manufacturer, monitored by the competent authority, must then perform an investigation to determine what exactly happened in the event. A format for an investigation report can be found in Annex 3 of the guidance document. Generally speaking, the manufacturer should report almost all events to the relevant competent authorities and the notified body. However, the guidance document specifies events that may not need to be reported, such as:
These events don’t trigger the manufacturer’s vigilance procedure, unlike an incident, and so don’t need to be reported. It should be noted that if any of these exceptions happen such that the manufacturer notices a trend, the manufacturer should report the trend to the competent authorities. The GHTF published a document for manufacturers that details the method a manufacturer could use to determine when a set of isolated events that wouldn’t otherwise be reportable should be reported.9 Vigilance in Action - Risk ManagementRisk management plays a large role in both the pre- and post-market phases Manufacturers must not only analyze the device’s risk before marketing, but must also continue to analyze the device’s risk throughout its lifetime. EN ISO 14971 provides clear rules for a manufacturer’s risk management file including the requirements and the process that manufacturers should maintain. Section three of EN ISO 14971:2009 lists the requirements for risk management:
Establishing a risk management process is important for manufacturers because it allows them to collect post-market clinical or non-clinical data on their device. Maintaining a risk management process allows a manufacturer to continually monitor the safety of their device and identify any new hazards or dangerous trends. To effectively manage a device’s risk, the manufacturer must also continually analyze and document the risk using the process outlined in EN ISO 14971:2009. In other words, as a part of the update of the risk management file, the PMCF study results is used to analyze the device’s risk throughout its lifetime. At this stage of the device’s life, the risk analysis should include, among others:
Assessing the Post-Market Surveillance System – A Notified Body’s RoleNotified Bodies step in at the start of the post-market surveillance timeline because it must assess a manufacturer’s post-market surveillance system as part of the manufacturer’s quality management system. A notified body, however, doesn’t help the manufacturer in its post-market surveillance duties. Instead, the notified body’s role is to verify that the manufacturer has a system in place, that the system conforms to the regulations and standards, and that the system works. To do this, the notified body first evaluates the planning of these activities. The notified body could take sample incidents or results of a PMCF study from the manufacturer’s files, audit them to ensure the system was followed, and confirm that the manufacturer implemented any corrective or preventive actions.10 When assessing a manufacturer’s vigilance system, a notified body first makes sure that the manufacturer’s system conforms to the regulations and standards. This means that the manufacturer defines which person or position will take on each role in the system and the timeframe to report any adverse events. The notified body also makes sure that the system is built in a way that the manufacturer could use it. For example, for a high-risk implantable device that must be traceable, the manufacturer can trace the device from its own factory through the distributor to the user and the end patient. On the other hand, a system that is not complete cannot give the manufacturer accurate data on the device and its safety and efficacy, which may prove problematic in the future. The notified body will also assess the method the manufacturer uses to collect and report incidents about their device. In this way, it is important for manufacturers to stay on top of the differences in the national laws for each country in which the manufacturer sells a device. Maintaining a post-market surveillance system that takes these differences into account is extremely important to ensure the manufacturer logs user complaints related to incidents properly according to the different countries. While having a functioning post-market surveillance system that is properly audited won’t prevent a device recall or adverse event, the system will ensure the patient’s safety isn’t compromised any more than it needs to be as it can be reasonably predictable using some very factual data. As a result, the post-market surveillance system is an important and integral part of the manufacturer’s quality management system. It is also part of the elements a company’s board of directors or CEO should take into account when defining the quality and strategic objectives, as it could impact the company’s public image. LNE/G-MED is a notified body with decades of experience in auditing and assessing quality management systems. For more information on how to use LNE/G-MED’s expertise in pointing out how your quality management system could improve, contact us.
1National Competent Authorities Summary Record of Notifications, 2003-2010, http://ec.europa.eu/health/medical-devices/files/summary_en.pdf; Nagreha, Nitin MD, Meeta Parmar. “Study of Medical Device Recalls by FDA Over Last Four Years (2008-2011).” Applied Clinical Trials Online. Sep. 20, 2011. 2MEDDEV 2.12-1 rev 6 3Guidelines on Medical Devices: Post Market Clinical Follow-Up Studies, A Guide for Manufacturers and Notified Bodies. MEDDEV 2.12/2 rev2. January 2012. 4Id. l.49-51. 5Id. l. 206-232. 6Id. l. 250-256. 7Id. l. 169 8MEDDEV 2.12-1 rev 6 §5.1.1 9GHTF/SG2/N36R7:2003 10NBOG 2009-2, Role of Notified Bodies in the Medical Device Vigilance System, §2.1
Get a Customized Proposal from our Technical Experts.Use our Certification Application for a detailed proposal. For more urgent requests, feel free to use our Quick Quote proposal form. For requests other than certification proposals, please use the following links:
|





- Progress Report - News
- MD Revisions Draft Amendments
- MD Classifications
- ISO 14971:2012
- MDD Revision: Actors, Roles and Responsabilities
- WEEE2 Directive
- Maintaining Compliance
- Revisions to the MDD
- ISO 13485 and FDA
- Controlling Outsourced Processes
- The Notified Body’s Role in Medical Device Certification
- EC Publishes E-Labeling Regulation
- IVD Updates
- Post-Market Phase
- IEC 60601 3rd Edition
- Tech File Evaluation
- Clinical Evaluation
- Recast of IVD Directive
- Own Brand Label Devices
- Software Engineering
- Med. Device Packaging
- Single Use Devices
- Usability Engineering
- FDA Use of ISO 13485
- Risk Management
- IVD Directive
- Switching Notified Body
- 60601 3rd Edition Status
- Regulatory News
- Ask Our Experts
- Free Guide: CE Marking
- Free Guide: IEC 60601 3rd Edition
- Free Guide: EMC Testing
- Free Guide: Brazil MD Approval
- Free Guide: Japan MD Approval
- Free Guide: IVD
- Technical FAQs
- Progress Report - News